Hailie Sensor

← Back to Device Overview

  • Data Transfer: Via Tenovi Gateway, whenever a reading is taken
  • Sensor Code: 25
  • On-Device Measurement Storage: Yes – measurements taken outside of range of the Tenovi Gateway will be saved locally and uploaded the next time the Gateway is in range.
  • Supports tracking multiple devices per gateway via Sensor IDs: Yes (required) – the Hailie Sensor’s Serial No must be included as the sensor_id when activating new devices. Measurements will be linked to the Client Devices with the Serial No of the physical device used.

Metrics

NameAPI FormatValue 1Value 2Is Vital?*
Hailie Medication Eventhailie_medication_event1N/A
Low Battery Eventlow_battery_event1 or 0 (critical/low)N/A
Hailie Turbohaler Rotation Eventhailie_turbohaler_rotation_event1 or 0 (click/non-click)N/A
Hailie Respiratory Flow Loghailie_respiratory_flow_logMax Flow Rate (L/M)Avg Flow Rate (L/M)
Hailie Respiratory Timing Loghailie_respiratory_timing_logDuration (s)Time to Peak (s)
Hailie Shake Detection Eventhailie_shake_detection_event1 or 0 (valid shake or not)Shake Duration (s)
Hailie Inhaler Installed Eventhailie_inhaler_installed_event1 or 0 (installed or removed)N/A
Hailie Medication Cap Eventhailie_medication_cap_event1 or 0 (cap on/cap off)N/A

Additional Data Included in Filter Params

KeyPossible ValuesDescription
angleNA0 to 22.5
22.5 to 45
45 to 67.5
67.5 to 90
90 to 112.5
112.5 to 135
135 to 157.5
157.5 to 180
The angle the device was in when the medication log was stored
flagskey-value pairsVariable vendor-defined flags (depending on metric type):hailie_respiratory_flow_log:inhalation_valid: 0/1flow_error: 0/1exhalation_detected: 0/1hailie_shake_detection_event:shake_intensity: positive integerhailie_medication_event or halie_turbohaler_rotation_event:respiration_enabled: 0/1orientation_enabled: 0/1orientation_stored: 0/1orientation_valid: 0/1 
serial_number6 to 8 alphanumeric charactersSerial number of the sensor associated with the event
device_log_indexAny Positive IntegerAn index value provided by the Hailie Sensor that can be used to identify if a reading is unique or a duplicate of a previous reading.

Additional Parameters That Can be Set Via Device Properties

KeyPossible ValuesDescription
measurement_webhook_idWebhook UUIDSet this property to configure a Device to only send measurement data to a single Webhook (specified by that Webhook’s ID).

Downloads & Device Support

DeviceResources
Adherium Hailie® Advair Diskus/Flovent DiskusQuick Start Guide
Adherium Hailie® Advair Flovent HFA (Next Gen)Quick Start Guide | Device Support
Adherium Hailie® Bevespi AerosphereQuick Start Guide
Adherium Hailie® Ellipta (Next Gen)Quick Start Guide | User Manual | Device Support
Adherium Hailie® HandihalerQuick Start Guide
Adherium Hailie® Symbicort HFA (Next Gen)Quick Start Guide | User Manual | Device Support
Adherium Hailie® Ventolin HFA (Next Gen)Quick Start Guide | Device Support
Adherium Hailie® for Teva ProAir® HFA (Next Gen)Quick Start Guide | Device Support

Last updated: